Accuracy of Optical Heart Rate Measurements for 10 Commercial Wearables in Different Climate Conditions and Activities: Instrument Validation Study

10款商用可穿戴设备在不同气候条件和活动下的光学心率测量准确性:仪器验证研究

阅读:1

Abstract

BACKGROUND: Commercial wearable devices allow for continuous heart rate (HR) monitoring in daily life. Their accuracy under ecologically valid conditions, however, remains insufficiently independently tested, especially during irregular activity, cognitive stress, and variable climates. OBJECTIVE: This study evaluated the HR accuracy of 10 commercially available wearables under controlled variations in physical activity, cognitive stress, and temperature. We hypothesized that physical activity irregularity, cognitive stress, and thermal climate conditions would affect measurement accuracy. METHODS: Forty-five healthy adults (21-68, mean 34, SD 12 y) completed a standardized protocol in climate-controlled chambers simulating neutral (23 °C), hot (36 °C), and cold (10 °C) conditions. Tasks included rest, cognitive stress (Montreal Imaging Stress Task), steady walking, and intermittent walking. Each of the 10 devices (Fitbit Charge 6, Fitbit Inspire 3, Garmin Vivosmart 5, Garmin Vivoactive 5, Apple Watch SE, Google Pixel Watch 2, Polar Ignite 3, Polar Pacer, Xiaomi Watch 2, and Oura Ring Gen 3) was compared against electrocardiogram-derived HR from a Zephyr BioHarness chest strap. Accuracy was assessed using mean absolute error (MAE), mean absolute percentage error (MAPE), repeated-measures concordance correlation coefficient (CCC), and Bland-Altman analysis. RESULTS: Significant variability across the devices was observed. Fitbit Charge 6 (MAE 4.5 bpm, MAPE 5.5%, CCC 0.93) and Google Pixel Watch 2 (MAE 4.9 bpm, MAPE 6.7%, CCC 0.87) showed strong agreement with the gold standard. In contrast, Fitbit Inspire 3, Polar Ignite 3, Polar Pacer, and Oura Ring displayed larger errors (MAE 9-14 bpm, MAPE 11%-16%) and lower CCC values (0.45-0.66). The climate conditions did not significantly affect the measurement accuracy of the test devices. The activity type, however, did have a significant effect: intermittent walking increased errors for multiple devices. CONCLUSIONS: Wearable HR measurement accuracy is device-specific and context-dependent. Moderate climates did not impair performance, but irregular movement reduced accuracy. Fitbit Charge 6 and Google Pixel Watch 2 demonstrated the highest reliability, supporting their use in health and sports monitoring. Careful device selection and context-aware interpretation remain critical for applied and clinical applications.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。