Abstract
Background: Stereotactic radiotherapy (SRT) is a promising treatment option for patients with multiple brain metastases (BMs). Using one isocenter instead of a separate isocenter for each BM can reduce the treatment time. This work compares the calculated dose in the treatment planning system with the measured dose using film dosimetry of single-isocenter multi-target (SIMT) SRT for multiple BM. Methods: Fifty patients with 4 to 18 BMs (median = 6, in total 356 BMs) were treated with a single-isocenter non-coplanar LINAC-based treatment with six VMAT arcs. Treatment was performed using RayStation and Elekta Versa HD with Agility multileaf collimator, including a 6D robotic couch. Patient-specific QA measurements were performed with an in-house developed phantom using three layers of GafChromic EBT3 film. Film measurements were analyzed in DoseLab using global gamma with 3% and 1 mm distance-to-agreement criteria. Additionally, secondary dose calculations in Mobius3D were performed with similar gamma criteria. Results: The mean total Paddick conformity index and gradient index were 0.7 ± 0.10 and 5.2 ± 1.9, respectively. Monitor units used were 6321 ± 2510, and mean irradiation time was 600 ± 90 s. The mean global gamma passing rate for all measured films was 94.5 ± 4.6% with 3% and 1 mm criteria, while that of the dose calculations in Mobius3D was 98.2 ± 1.2% with the same criteria. A dependence of gamma passing rates of film measurements on the total PTV volume was observed, whereas such dependence was minimal for Mobius3D. Conclusions: The results demonstrate good agreement between the TPS, film measurements, and independent dose calculations, supporting the dosimetric accuracy of single-isocenter multi-target SRT for treating multiple BMs.