Establishment of a Novel Method for Dose Distribution Verification in Stereotactic Radiosurgery (SRS) Using General Semiconductor Array Detectors

利用通用半导体阵列探测器建立立体定向放射外科(SRS)剂量分布验证的新方法

阅读:2

Abstract

Stereotactic radiosurgery (SRS) involves steep dose gradients and small irradiation fields, which makes ensuring irradiation accuracy extremely important. While general semiconductor array detectors are reusable and cost-effective, their wide detector spacing results in insufficient spatial resolution. The aim of this study is to establish a novel patient-specific quality assurance (PSQA) method for single-isocenter multiple-target SRS volumetric modulated arc therapy (SIMT-SRS-VMAT) that can evaluate dose distribution with the same accuracy as radiochromic film by virtually narrowing the detector spacing by combining measurement with a general semiconductor array detector and translational movement of the treatment couch. Single-field plans with a single target and SIMT-SRS-VMAT plans with two targets were created for 1 cm lesions. Measurements using MapCHECK2 (Sun Nuclear, Melbourne, FL, USA) were combined with couch movement to synthesize measurement data at virtual detector spacings of 5, 4, 3, 2, and 1 mm. Root mean square error (RMSE) was used to evaluate profile shape accuracy, and γ analysis was used to evaluate dose distribution agreement. At virtual detector spacings of 10 mm, 5 mm, 4 mm, 3 mm, 2 mm, 1 mm, and with radiochromic film, the RMSE values in SIMT-SRS-VMAT were 8.02% ± 0.34%, 2.16% ± 0.07%, 1.60% ± 0.08%, 1.48% ± 0.17%, 1.32% ± 0.10%, 1.38% ± 0.03%, and 1.25% ± 0.02%, respectively. The RMSE decreased as the detector spacing narrowed, resulting in profile shapes comparable to those obtained with radiochromic film. In γ analysis, setting the virtual detector spacing to 5 mm enabled analysis covering the entire dose range, and the γ pass rate was comparable to EBT4 (Ashland Inc., Wayne, NJ, USA) for all tolerances. This study established a novel method of verification of dose distribution with the same accuracy as film, combining measurement with a general semiconductor array detector and translational movement of the treatment couch. With this approach, specialized high-resolution SRS array detectors and disposable radiochromic film are not needed, making SRS PSQA less costly and manpower-intensive.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。