Initial three- and half-year machine performance assessment of a superficial radiation therapy unit

对一台浅表放射治疗机进行为期三年半的初步机器性能评估

阅读:1

Abstract

PURPOSE: This work reports on results of periodic machine quality assurance (QA) tests for our superficial radiation therapy (SRT) unit (SRT-100 Vision) performed for the first three- and half- year. METHODS: Results of machine warm-up, dosimetry, mechanical, safety and imaging QA tests for our SRT-100 Vision unit were reviewed and analyzed. Dosimetry tests included output constancy, backup timer, timer accuracy, beam quality (half value layer [HVL]), applicator factors (AFs), absolute output calibration, output constancy with varying x-ray tube head rotation, output reproducibility, output linearity, timer and end-effect error, percent depth dose (PDD) verification and congruence between radiation field and applicator size. Mechanical tests encompassed unit stability as well as applicator integrity and indicators. Safety tests consisted of visual and audio monitors, beam-on indicator, interlocks, and energy/filter indicators. Imaging tests for ultrasound covered functional check, spatial integrity, and image quality check. The QA tests were performed for three kilovoltage (kV) x-ray energies (50, 70, and 100 kV(p)) and six applicators (1.5, 2.0, 2.5, 3.0, 4.0, and 5.0 cm in diameter). RESULTS: Daily mandatory machine warm-up was successfully completed on each day of treatment. The results of all dosimetry tests for three energies were within the recommended tolerance. Daily, and monthly outputs and annual absolute outputs were < ± 3.0%, < ± 2.0%, and < ± 1.3%, respectively. Backup timer displayed timer set ± 0.0%. Timer accuracy was < ± 1.0 sec. HVLs and AFs were < ± 1.1% and < ± 0.3%, respectively. Outputs for different head angles were < ± 1.0% and output reproducibility was within 0.1%. Output linearity was < ± 0.6% and end-effect errors were < ± 0.0006 min. PDDs were < ± 2.2%. Congruence was < ± 0.5 mm. Monthly mechanical tests confirmed mechanical robustness of our unit. Daily safety features were verified to be functional each day. Monthly imaging tests for ultrasound showed good functionality, image quality, and spatial integrity meeting the set tolerance. CONCLUSIONS: The QA results demonstrated high performance and stability of our SRT-100 Vision unit over the first three and a half years of operation.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。