Feasibility of HyperSight CBCT for adaptive radiation therapy: A phantom benchmark study of dose calculation accuracy and delivery verification on the Halcyon

HyperSight CBCT在自适应放射治疗中的可行性:基于Halcyon的剂量计算精度和实施验证的体模基准研究

阅读:1

Abstract

PURPOSE: The development of on-board cone-beam computed tomography (CBCT) has led to improved target localization and evaluation of patient anatomical change throughout the course of radiation therapy. HyperSight, a newly developed on-board CBCT platform by Varian, has been shown to improve image quality and HU fidelity relative to conventional CBCT. The purpose of this study is to benchmark the dose calculation accuracy of Varian's HyperSight cone-beam computed tomography (CBCT) on the Halcyon platform relative to fan-beam CT-based dose calculations and to perform end-to-end testing of HyperSight CBCT-only based treatment planning. METHODS: A HU to mass density curve was measured for the HyperSight CBCT system and implemented into the Eclipse treatment planning system. Following this, computational dosimetric analysis was performed between dose distributions calculated on CT simulation (CT(sim)) and HyperSight CBCT images on two anthropomorphic phantoms for pelvic and head and neck treatment sites. Additionally, an end-to-end test was carried out for a head and neck intensity modulated radiation therapy (IMRT) plan. RESULTS: The HU to mass density curves acquired on CT(sim) and HyperSight CBCT were similar (< 30 HU) for near-water equivalent materials, but deviated for high-density materials, with a maximum difference of 150 HU. For dose calculations, excellent agreement between dose calculations performed on CT(sim) and HyperSight CBCT phantom images was observed, where three-dimensional gamma pass rates between the two dose distributions were observed to be  ≥  90% at 1%/1 mm (5% threshold). For the end-to-end test, absolute doses were verified to within 1% of ionization chamber measurements, while Delta4+ and portal dosimetry measurements yielded passing results (gamma pass rate  ≥  90%) down to 2%/2 mm criterion. CONCLUSION: In this study, the accuracy of dose calculations performed on HyperSight CBCT was found to be within 1% of CT(sim) calculations for pelvic and head and neck treatment sites. End-to-end results using the RANDO anthropomorphic phantom indicate that HyperSight CBCT images are suitable for radiation treatment planning.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。