Comparative Evaluation of Indirect Immunofluorescence and NS-1-Based ELISA to Determine Zika Virus-Specific IgM

比较间接免疫荧光法和基于NS-1的ELISA法测定寨卡病毒特异性IgM的效能

阅读:1

Abstract

Differential diagnosis of the Zika virus (ZIKV) is hampered by cross-reactivity with other flaviviruses, mainly dengue viruses. The aim of this study was to compare two commercial methods for detecting ZIKV immunoglobulin M (IgM), an indirect immunofluorescence (IIF) and an enzyme immunoassay (ELISA), using the non-structural (NS) 1 protein as an antigen, both from EuroImmun, Germany. In total, 255 serum samples were analyzed, 203 of which showed laboratory markers of ZIKV infections (PCR-positive in serum and/or in urine and/or positive or indeterminate specific IgM). When tested with IIF, 163 samples were IgM-positive, while 13 samples were indeterminate and 78 were negative. When IIF-positive samples were tested using ELISA, we found 61 positive results, 14 indeterminate results, and 88 negative results. Among the indeterminate cases tested with IIF, ELISA analysis found two positive, two indeterminate, and nine negative results. Finally, 74 of the 78 IIF-negative samples proved also to be negative using ELISA. For the calculations, all indeterminate results were considered to be positive. The agreement, sensitivity, and specificity between ELISA and IIF were 60.2%, 44.9%, and 94.9%, respectively. Overall, 101 samples showed discrepant results; these samples were finally classified on the basis of other ZIKV diagnostic approaches (PCR-positive in serum and/or in urine, IgG determinations using IIF or ELISA, and ZIKV Plaque Reduction Neutralization test-positive), when available. A final classification of 228 samples was possible; 126 of them were positive and 102 were negative. The corresponding values of agreement, sensitivity, and specificity of IIF were 86.0%, 96.8%, and 72.5%, respectively. The corresponding figures for ELISA were 81.1%, 65.9%, and 100%, respectively. The ELISA and IIF methods are both adequate approaches for detecting ZIKV-specific IgM. However, considering their respective weaknesses (low sensitivity in ELISA and low specificity in IIF), serological results must be considered jointly with other laboratory results.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。