Prospective evaluation of an ultrathin strut biodegradable polymer-coated sirolimus-eluting stent: 12 months' results from the S-FLEX UK registry

前瞻性评估超薄支架可生物降解聚合物涂层西罗莫司洗脱支架:来自S-FLEX英国注册研究的12个月结果

阅读:1

Abstract

OBJECTIVE: To prospectively evaluate safety and efficacy of the ultrathin strut biodegradable polymer-coated Supraflex sirolimus-eluting stent (S-SES) in 'real world' patient population requiring percutaneous coronary intervention (PCI). METHODS: National, prospective, multicentre, single-arm, all-comers, observational registry of 469 patients treated with S-SES from July 2015 and November 2016 in 11 centres in UK. Primary endpoint was target lesion failure (TLF) at 12 months (cardiac death, target vessel myocardial infarction (MI) or clinically driven target lesion revascularisation (TLR)). Secondary endpoints included safety and performance outcomes at 12 months-overall stent thrombosis (ST), all-cause mortality, any MI, target vessel failure (TVF) and major adverse cardiac events (MACE-composite of cardiac death, MI, emergent or repeat revascularisation). RESULTS: At 12 months, the primary endpoint occurred in 11 (2.4%) of 466 patients, consisting of 4 (0.9%) cardiac deaths, 3 (0.6%) target vessel MI and 7 (1.5%) TLR. Secondary endpoints findings included all-cause mortality in 6 (1.3%), TVF of 14 (3%), no definite ST, 1 (0.2%) probable ST and 3 (0.6%) possible ST. Overall MACE was observed in 18 (3.9%). CONCLUSIONS: The S-FLEX UK registry showed that the S-SES is safe with a low incidence of TLF in routine clinical practise in patients with coronary artery disease being treated by PCI.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。