Noise Produced by Neonatal Ventilators Inside and Outside of the Incubators

新生儿呼吸机在保温箱内外产生的噪音

阅读:1

Abstract

BACKGROUND: Insufficient data are available about the noise produced by modern neonatal ventilators. We aimed to measure their noise under different ventilatory modes and parameters. METHODS: This was a bench study measuring the noise produced by 9 neonatal ventilators set in conventional or high-frequency oscillatory ventilation (HFOV), nasal mask-delivered CPAP with variable- or continuous-flow configuration, or bi-level positive airway pressure (considered as noninvasive ventilation [NIV]). Conventional ventilation and HFOV were tested in 2 distinct settings with moderate or higher parameters. Sound measurements were performed inside and outside an incubator mimicking the clinical setting and using a high-end meter meeting the international ISO 226:2003 standard. RESULTS: Four ventilators remained below the internationally recommended safety threshold but only for measurements outside the incubator. Conventional ventilation (49.1 [3.4] dBA) and HFOV (56.3 [5.2] dBA) were the least and most noisy respiratory support technique, respectively. Noise was greater inside than outside the incubators (P < .0001) and different between the ventilators (P < .0001); better results were achieved by Servo-u and Fabian family devices for conventional ventilation; by fabian HFO for HFOV; and by Servo-u, VN500, and fabian family devices for CPAP and NIV. Noise levels were similar when using moderate or higher parameters in conventional ventilation (P = .81) and in HFOV (P = .45). CONCLUSIONS: Modern ventilators often produce relevant noise, independent of the respiratory support modality, with acceptable noise levels being measured only outside the incubator. Better results were achieved with Servo-u, VN500, and Fabian family devices.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。