Accuracy of Multiple Pulse Oximeters in Stable Critically Ill Patients

多台脉搏血氧仪在病情稳定的危重患者中的准确性

阅读:1

Abstract

BACKGROUND: An accurate S(pO(2)) value is critical in order to optimally titrate oxygen delivery to patients and to follow oxygenation guidelines. Limited prospective data exist on real-world performance of pulse oximeters in critically ill patients. The objective of this study was to assess accuracy and bias of the S(pO(2)) values measured by several oximeters in hospitalized subjects. METHODS: We included stable adults in the ICU with an arterial catheter in place. Main exclusion criteria were poor S(pO(2)) signal and S(pO(2)) > 96%. In each subject, we simultaneously evaluated 4 oximeters: Nonin (Plymouth, Minnesota) embedded in the FreeO(2) device (OxyNov, Québec City, Québec, Canada), Masimo (Radical-7, Masimo, Irvine, California), Philips (FAST, Philips, Amsterdam, the Netherlands), and Nellcor (N-600, Medtronic, Minneapolis, Minnesota). Arterial blood gases were drawn and simultaneously each oximeters' S(pO(2)) values were collected. S(pO(2)) values were compared to the reference (arterial oxygen saturation [S(aO(2)) ] value) to determine bias and accuracy. The ability for oximeters to detect hypoxemia and the impact of oximeters on oxygen titration were evaluated. RESULTS: We included 193 subjects (153 male, mean age 66 y) in whom 211 sets of measurements were performed. The skin pigmentation evaluated by Fitzpatrick scale showed 96.2% of subjects were light skin (types 1 and 2). One oximeter overestimated S(aO(2)) (Philips, +0.9%), whereas the 3 others underestimated S(aO(2)) (Nonin -3.1%, Nellcor -0.3%, Masimo -0.2%). S(aO(2)) was underestimated with Nonin oximeter in 91.3% of the cases, whereas it was overestimated in 55.2% of the cases with Philips oximeter. Moderate hypoxemia (S(aO(2)) 86-90% or P(aO(2)) 55-60 mm Hg) was detected in 92, 33, 42, and 11% of the cases with Nonin, Nellcor, Masimo, and Philips, respectively. CONCLUSIONS: We found significant bias and moderate accuracy between the tested oximeters and the arterial blood gases in the studied population. These discrepancies may have important clinical impact on the detection of hypoxemia and management of oxygen therapy.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。