Comparison of clinical outcomes following delivery of budesonide by both vibrating mesh nebulizer and jet nebulizer in premature infants with bronchopulmonary dysplasia

比较振动网式雾化器和喷射式雾化器治疗早产儿支气管肺发育不良后布地奈德的临床疗效

阅读:1

Abstract

BACKGROUND: This study aimed to compare the efficacy of budesonide inhalation suspension administered via a vibrating mesh nebulizer vs. a jet nebulizer in the treatment of premature infants with bronchopulmonary dysplasia (BPD) undergoing high-frequency oscillatory ventilation (HFOV). METHODS: Between July 2020 and July 2022, we retrospectively analyzed the medical records of 36 preterm infants diagnosed with BPD who underwent HFOV. Based on the nebulizer type used, infants were categorized into the vibrating mesh nebulizer group (VMN group) or the jet nebulizer group (JN group). Post-nebulization outcomes, such as the duration of mechanical ventilation, length of stay in the neonatal intensive care unit (NICU), ventilator-associated parameters, and arterial blood gas metrics, were compared between the two groups. Treatment-associated complications were also documented. RESULTS: No significant differences were noted between the VMN and JN groups in terms of mechanical ventilation duration (p = 0.519), NICU length of stay (p = 0.112), ventilator-associated parameters, or complications (p = 0.700). However, after 2 weeks of treatment, the oxygenation index (p = 0.012) and arterial partial pressure of carbon dioxide (p = 0.006) were more favorable in the VMN group compared to the JN group. CONCLUSION: Among premature infants with BPD on HFOV, for administration of budesonide inhalation suspension resulted in an improved oxygenation index and reduced arterial partial pressure of carbon dioxide when compared to a jet nebulizer, indicating superior therapeutic efficacy.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。