Extracorporeal membrane oxygenation and COVID-19: The causes of failure

体外膜肺氧合(ECMO)与新冠肺炎:失败的原因

阅读:1

Abstract

INTRODUCTION: Venovenous extracorporeal membrane oxygenation (VV-ECMO) is a therapeutic strategy for the coronavirus disease 2019 (COVID-19) induced acute respiratory distress syndrome (ARDS). There are inconclusive data in this regard and causes of VV-ECMO failure are not yet understood well. CASE SERIES: Here, seven patients with COVID-19-induced ARDS who underwent VV-ECMO introduced and causes of VV-ECMO failure discussed. Medical records of seven COVID-19 patients treated with VV-ECMO were retrospectively evaluated to determine the clinical outcomes of VV-ECMO. Oxygenator failure occurred in four patients whom needed to oxygenator replacement. Successful VV-ECMO decannulation was done in three patients, however finally one patient survived. CONCLUSIONS: Hypercoagulability state and oxygenator failure were the most main etiologies for VV-ECMO failure in our study. All patients with COVID-19 undergoing VV-ECMO should be monitored for such problems and highly specialized healthcare team should monitor the patients during VV-ECMO.