Evaluation of the Humidification Performances of New Generation of Heated Wire Humidifiers

新一代加热丝加湿器加湿性能评价

阅读:1

Abstract

BACKGROUND: At high ambient temperatures in ICU rooms, the humidification performances of heated-wire humidifiers are significantly reduced, with delivered gas humidity well below 30 mg H(2)O/L, which leads to an increased risk of endotracheal occlusions, subocclusions, or mucociliary dysfunction. The objective of the study was to evaluate the humidity delivered at the Y-piece with new-generation heated-wire humidifiers with advanced algorithm (FP950 [Fisher & Paykel Healthcare, Auckland, New Zealand] and VHB20 [Vincent Medical, Inspired, Hong Kong]) while varying ambient temperatures. METHODS: We measured, on the bench, the hygrometry of inspiratory gases delivered by a new generation of heated-wire humidifiers (i) FP950, (ii) VHB20 and a previous generation of heated-wire humidifiers, (iii) MR850 (Fisher & Paykel) with the usual settings (37°C at the chamber/40°C at the Y-piece), (iv) MR850 with no temperature gradient (40°C/40°C), and (v) MR850 with the automatic compensation algorithm activated. Hygrometry was measured with the psychrometric method after 1 h of stability while varying the room temperature from 20 to 30°C. RESULTS: Two hundred ninety-four hygrometric bench measurements were performed at steady state for the different tested conditions. With the new heated-wire humidifiers (FP950 and VHB20), gas humidity delivered remained > 30 mg H(2)O/L in all tested conditions, even at high ambient temperatures (>25°C). With previous generations of heated-wire humidifiers (MR850), at high ambient temperature, humidity delivered was adequate in only 26% (11/42) of the measurements when the usual settings were used (37°C/40°C) and 30% (11/37) with automatic compensation. When no temperature gradient was set (40°C/40°C), humidity delivered was > 30 mg H(2)O/L in 91% (30/33) of the measurements at a high ambient temperature. With an ambient temperature < 25°C, almost all devices and settings provided adequate humidity. CONCLUSIONS: The new FP950 and VHB20 heated-wire humidifiers by using advanced algorithms demonstrated stable performance while varying the ambient temperature by 20-30°C, better than the previous generation of heated humidifiers when ambient temperatures were high.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。