Validating the accuracy of the Omron HBP-M4500 blood pressure measuring device according to the ISO81060-2:2018+ Amd 1:2020 protocol

根据 ISO81060-2:2018+ Amd 1:2020 协议验证欧姆龙 HBP-M4500 血压测量仪的准确性

阅读:2

Abstract

BACKGROUND: Blood pressure (BP) naturally undergoes fluctuations and variations, particularly during anesthesia administration during surgery, hemodialysis, upper and lower gastrointestinal endoscopy, exercise testing, arterial and venous catheterization, and rehabilitation. These changes in BP may lead to life-threatening events. OBJECTIVES: The performance of the Omron HBP-M4500 device in monitoring blood pressure (BP) in the upper arm was validated according to the International Organization for Standardization (ISO) 81060-2:2018+amendment (Amd) 1:2020 protocol. METHODS: The device was used to assess 113 participants in the inflation mode, and 107 participants in the deflation mode. All the patients fulfilled the inclusion criteria, including the arm circumference range and systolic and diastolic BP levels, outlined in the protocol. Data validation and analysis were performed according to the manufacturer's instructions. RESULTS: In criterion 1, the mean ± standard deviation (SD) values of the differences between the test device and reference BP were -0.6 ± 5.80/2.8 ± 6.78 mmHg (systolic/diastolic) and -1.0 ± 5.35/3.2 ± 6.52 mmHg for the inflation and deflation modes, respectively. These data fulfilled the ISO81060-2:2018+Amd1:2020 requirements of ⩽ 5 ±⁣⩽ 8 mmHg. In criterion 2, the differences were -0.6 ± 4.44/2.8 ± 6.26 and -1.0 ± 3.84/3.2 ± 6.09 mmHg for the inflation and deflation modes, respectively, fulfilling criterion 2 with SD values of ⩽ 6.91 and ⩽ 6.87 for systolic BP and ⩽ 6.34 and ⩽ 6.14 for diastolic BP in the inflation and deflation modes, respectively. These two criteria were fulfilled in both studies. CONCLUSION: The Omron HBP-M4500 device, either in inflation or deflation mode, fulfilled the criteria outlined in the ISO protocol. Therefore, this device is valuable for BP measurement in clinical and hospital settings.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。