Validation of the AViTA upper-arm blood pressure monitor BP636 with multiple cuff sizes in the general population according to the ANSI/AAMI/ISO 81060-2:2018/Amd1:2020

根据 ANSI/AAMI/ISO 81060-2:2018/Amd1:2020 标准,对 AViTA 上臂血压监测仪 BP636 在一般人群中采用多种袖带尺寸进行验证

阅读:4

Abstract

OBJECTIVE: To determine the accuracy of the AViTA BP636 upper-arm home blood pressure (BP) monitor in general populations with small to extra-large arms according to the ANSI/AAMI/ISO Universal Standard (ISO 81060-2 : 2018/Amd1 : 2020). METHODS: The AViTA BPM636 was utilized on 85 adult participants with upper-arm circumferences ranging from 15 to 52 cm, and four cuffs of different sizes (small: 15-22 cm, normal: 22-33 cm, large: 33-42 cm, and extra large: 42-52 cm) were used in this study. The recruitment fulfills the criteria of the ANSI/AAMI/ISO Universal Standard 81060-2 : 2018/Amd1 : 2020 in the general population using the same arm sequential BP measurement method. RESULTS: For validation criterion 1, the mean ± SD of the differences between the test device and reference blood pressure readings was -0.35 ± 4.68/0.56 ± 3.93 mmHg (systolic/diastolic) for adults. The mean difference of SBP between the test device and the reference device was -0.35 mmHg with a SD of 3.50 mmHg, which was within the acceptable range for criterion 2 (i.e. less than 6.93 mmHg) Similarly, the mean difference in DBP was 0.56 mmHg with a SD of 3.38 mmHg, which also met the criterion 2 (i.e. within 6.91 mmHg). CONCLUSION: AViTA BPM636 passed the criteria of the ANSI/AAMI/ISO 81060-2 : 2018/Amd 1 : 2020 protocol. It can be recommended for home-based BP measurements in the general population and people with extra-large arms ranging from 42 to 52 cm.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。