Direct comparison of an automated oscillometric device with an electronic auscultatory device for epidemiologic survey to evaluate the prevalence of hypertension

直接比较自动示波法血压计和电子听诊法血压计在流行病学调查中评估高血压患病率的应用

阅读:1

Abstract

Mercury-free sphygmomanometers are gradually replacing the traditional sphygmomanometers in clinical settings and epidemiological surveys for measuring blood pressure (BP) due to mercury toxicity. No direct comparative studies have evaluated BP differences and statistical errors of automated oscillometric devices (ODs) against electronic auscultatory devices (ADs) for epidemiologic surveys. Herein, we evaluated the validity of ODs for the Korea National Health and Nutrition Examination Survey (KNHANES) using the Universal Standard for BP device validation through a direct comparison with ADs as the reference standard. Four trained observers performed validation on 278 volunteers aged ≥ 19 years with a standardized BP measurement protocol. Agreement between the BP measurements recorded with an OD against those recorded with an AD was assessed by Lin's concordance correlation coefficient (CCC) and Bland-Altman's limits of agreement. To evaluate the agreement for BP classification, weighted kappa values were estimated. To explore the factors associated with BP measurement differences between the 2 devices, multiple linear regression analysis was performed. The average BP differences (OD-AD) were 2.6 ± 6.2 mm Hg for systolic BP (SBP) and -5.1 ± 5.6 mm Hg for diastolic BP (DBP). Lin's CCCs were 0.927 and 0.768 for the overall SBP and DBP, respectively. The cumulative percentage of absolute errors ≤10 mm Hg was 88.1% for SBP and 81.3% for DBP. The weighted kappa value for the Joint National Committee 7 BP classification was 0.75 (95% confidence interval: 0.68-0.81). An OD overestimated the prevalence of SBP (0.3%, P = .0222) and underestimated the prevalence of DBP (1.8%, P < .0001). Multivariate analysis to identify the risk factors for BP difference revealed the arm circumference (AC) to be negatively associated with BP difference. Male sex was positively associated, while age was negatively associated with SBP difference. OD-DBP was positively associated with DBP difference and negatively associated for DBP absolute error. ODs met the accuracy requirements of the Universal Standard criteria against ADs for SBP but not for DBP. Thus, the DBP values may be underestimated by ODs in the KNHANES.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。