Organ and effective dose reduction in adult chest CT using abdominal lead shielding

使用腹部铅屏蔽降低成人胸部CT的器官和有效剂量

阅读:2

Abstract

OBJECTIVES: The purpose of this study was to evaluate and compare organ and effective dose savings that could be achieved using conventional lead aprons and a new, custom-designed shield as out-of-plane shielding devices during chest CT scans. METHODS: Thermoluminescent dosimeters were used to measure doses throughout the abdomen and pelvis during CT scans of the chest of a RANDO phantom. Dose measurements were made with no shielding, with lead aprons and with the new shield around the abdomen and pelvis in order to quantify the achievable organ and effective dose reductions. RESULTS: Average dose savings in the 10 phantom sections ranged from 5% to 78% with the highest point dose saving of 93% being found in the mid-pelvis. When shielding was used, the maximum measured organ dose reduction was a 72% dose saving to the testes. Significant dose savings were found throughout the abdomen and pelvis, which contributed to an effective dose saving of 4% that was achieved over and above the dose savings obtained through conventional optimisation strategies. This could yield significant population dose savings and reductions in collective radiation risk. CONCLUSION: In this study significant organ and effective dose reductions have been achieved through the use of abdominal shielding during chest CT examinations and it is therefore recommended that out-of-plane patient shielding devices should be used for all chest CT scans and potentially for every CT scan, irrespective of body part.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。