Radiation dose reduction in chest dual-energy computed tomography: effect on image quality and diagnostic information

胸部双能计算机断层扫描辐射剂量降低:对图像质量和诊断信息的影响

阅读:1

Abstract

OBJECTIVE: To determine whether dual-energy computed tomography (DECT) of the chest can be performed at a reduced radiation dose, with an emphasis on images generated with post-processing techniques. MATERIALS AND METHODS: In 21 patients undergoing DECT of the chest in a dual-source scanner, an additional image series was acquired at a reduced radiation dose. Four thoracic radiologists assessed both image series for image quality, normal thoracic structures, as well as pulmonary and mediastinal abnormalities, on virtual monochromatic images at 40 keV and 60 keV. Data were analyzed with Student's t-test, kappa statistics, analysis of variance, and the Wilcoxon signed-rank test. RESULTS: The overall image quality of 60 keV virtual monochromatic images at a reduced radiation dose was considered optimal in all patients, and no abnormalities were missed. Contrast enhancement and lesion detection performance were comparable between reduced-dose images at 40 keV and standard-of-care images at 60 keV. The intraobserver and interobserver agreement were both good. The mean volumetric CT dose index (CTDIvol), size-specific dose estimate (SSDE), dose-length product (DLP), and effective dose (ED) for reduced-dose DECT were 3.0 ± 0.6 mGy, 4.0 ± 0.6 mGy, 107 ± 30 mGy.cm, and 1.5 ± 0.4 mSv, respectively. CONCLUSION: DECT of the chest can be performed at a reduced radiation dose (CTDIvol < 3 mGy) without loss of diagnostic information.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。