Acoustic response and ambient pressure sensitivity characterization of SonoVue for noninvasive pressure estimation

SonoVue无创压力估计的声学响应和环境压力灵敏度表征

阅读:1

Abstract

Subharmonic aided pressure estimation (SHAPE) is a noninvasive pressure measurement technique based on the pressure dependent subharmonic signal from contrast microbubbles. Here, SonoVue microbubble with a sulfur hexafluoride (SF6) core, was investigated for use in SHAPE. The study uses excitations of 25-700 kPa peak negative pressure (PNP) and 3 MHz frequency over eight pressurization cycles between atmospheric pressure and overpressures, ranging from 0 to 25 kPa (0 to 186 mm Hg). The SonoVue subharmonic response was characterized into two types. Unlike other microbubbles, SonoVue showed significant subharmonic signals at low excitations (PNPs, 25-400 kPa), denoted here as type I subharmonic. It linearly decreased with increasing overpressure (-0.52 dB/kPa at 100 kPa PNP). However, over multiple pressurization-depressurization cycles, type I subharmonic changed; its value at atmospheric pressure decreased over multiple cycles, and at later cycles, it recorded an increase in amplitude with overpressure (highest, +13 dB at 50 kPa PNP and 10 kPa overpressure). The subharmonic at higher excitations (PNP > 400 kPa), denoted here as type II subharmonic, showed a consistent decrease with the ambient pressure increase with strongest sensitivity of -0.4 dB/kPa at 500 kPa PNP.