UiO-66 Selective Enrichment Integrated with Thermal Desorption GC-MS for Detection of Benzene Homologues in Ambient Air

UiO-66选择性富集与热脱附GC-MS联用技术用于检测环境空气中的苯同系物

阅读:1

Abstract

In this study, UiO-66 was selected as sorbent media packed in the tube to selectively enrich trace levels of benzene homologues such as benzene, toluene, and xylene (BTX) in ambient air prior to thermal desorption (TD)-GC-MS determination. A series of experiments were conducted to obtain the optimal TD conditions. The results indicated that the optimal TD parameters were as follows: desorption temperature of 180°C, desorption flow rate of 50 mL min(-1), and desorption time of 30 min. Furthermore, the method based on UiO-66 enrichment integrated with TD-GC-MS for trace levels of BTX was successfully developed. It exhibited a good linearity (R (2) > 0.99) in the range of 50-1000 ng, except for p, m-xylene in the range of 100-2000 ng, and achieved the recovery of 69.4-101.3%, and the relative standard deviation of 3.8-6.4%. The detection limits of BTX were 1.6-4.0 ng; according to 10 L of sampling volume, the method detection limits would be in the range of 0.16-0.40 µg m(-3). Additionally, the method was successfully applied to determine BTX in indoor air and showed good selectivity and sensitivity. In summary, the findings in this work revealed that UiO-66 was an attractive adsorbent for selective enrichment trace levels of BTX compounds in ambient air, which was favorable for the subsequent detection by TD-GC-MS.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。