Development of a thoron calibration chamber based on computational fluid dynamics simulation and validation with measurements

基于计算流体动力学模拟和测量验证的钍校准室的研制

阅读:1

Abstract

Recently, interest in measuring the concentration of (220)Rn in air has increased greatly following the development of standards and the calibration of monitoring instruments. In this study, a (220)Rn calibration chamber was designed and developed at the Institute of Radiochemistry and Radioecology (RRI) based on the computational fluid dynamics (CFD) method implemented in ANSYS Fluent 2020 R1 code at the University of Pannonia in Hungary. The behavior of (220)Rn and its spatial distribution inside the (220)Rn calibration chamber at RRI were investigated at different flow rates. The (220)Rn concentration was close to homogeneous under higher flow regimes due to thorough mixing of the gas inside the chamber. Predictions based on CFD simulations were compared with experimentally measured transmission factors (C(out)/C(in)). The spatial distribution of (220)Rn was dependent on the flow rate and the positions of the inlet and outlet. Our results clearly demonstrate the suitability of the (220)Rn calibration chamber at RRI for calibrating monitoring instruments. Furthermore, the CFD-based predictions were in good agreement with the results obtained at higher flow rates using experimental and analytical models according to the relative deviation, with a maximum of approximately 9%.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。