Validation of an Intra-Oral Scan Method Versus Cone Beam Computed Tomography Superimposition to Assess the Accuracy between Planned and Achieved Dental Implants: A Randomized In Vitro Study

验证口内扫描方法与锥形束计算机断层扫描叠加法在评估计划种植牙与实际种植牙准确性方面的有效性:一项随机体外研究

阅读:1

Abstract

Computer aided implantology is the safest way to perform dental implants. The research of high accuracy represents a daily effort. The validated method to assess the accuracy of placed dental implants is the superimposition of a pre-operative and a post-operative cone beam computed tomography (CBCT) with planned and placed implants. This procedure is accountable for a biologic cost for the patient. To investigate alternative procedure for accuracy assessment, fifteen resin casts were printed. For each model, six implants were digitally planned and then placed following three different approaches: (a) template guided free hand, (b) static computer aided implantology (SCAI), and (c) dynamic computer aided implantology (DCAI). The placement accuracy of each implant was performed via two methods: the CBCT comparison described above and a matching between implant positions recovered from the original surgical plan with those obtained with a post-operative intraoral scan (IOS). Statistically significant mean differences between guided groups (SCAI and DCAI) and the free hand group were found at all considered deviations, while no differences resulted between the SCAI and DCAI approaches. Moreover, no mean statistically significant differences were found between CBCT and IOS assessment, confirming the validity of this new method.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。