Clinical Validation of a Highly Sensitive GC-MS Platform for Routine Urine Drug Screening and Real-Time Reporting of up to 212 Drugs

临床验证高灵敏度GC-MS平台在常规尿液药物筛查和实时报告多达212种药物方面的应用

阅读:1

Abstract

An important role of the clinical toxicology laboratory is to provide continuous diagnostic testing for patients with altered mental status and for other medical indications. To meet these needs, we have developed a new Gas Chromatography-Mass Spectrometry (GC-MS) platform that facilitates routine screening and automated reporting of 212 drugs by laboratory technologists around the clock without the need to sign out by an on-site mass spectrometry-trained toxicologist. The platform uses a programmable temperature vaporizer (PTV) injector for large sample volume injection and the free software Automated Mass Spectral Deconvolution and Identification System (AMDIS) for data reduction and spectral matching that facilitates rapid library searching and analyte identification. Method comparison with 118 patient samples demonstrated that this platform and data searching algorithm independently provided improvements in sensitivity compared to an established GC-MS platform. Further examination of the role of the data processing software and the in-house databases used in the established versus the new platform demonstrated that the improved analytical sensitivity of the new platform was attributed to both the technical superiority of the new GC-MS instrumentation and the use of AMDIS in conjunction with the newly generated in-house library for data processing.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。