Effects of light scattering on optical-resolution photoacoustic microscopy

光散射对光学分辨率光声显微镜的影响

阅读:1

Abstract

The imaging depth of ballistic optical imaging technologies is limited by light scattering. To study the effects of scattering on optical-resolution photoacoustic microscopy (OR-PAM), the signals were divided into target and background signals. A method to simulate the point spread function (PSF) of the PAM system considering both optical illumination and acoustic detection was proposed, then the PSF was used to calculate the contribution of each class of signal at different depths of the focal plane (zf). How image contrast is degraded when there is a uniformly absorbing background as well as when there are small targets densely packed in the acoustic resolution cell were studied. By using the hyperboloid-focusing-based Monte Carlo method, optical focusing into a scattering medium was simulated. It was found that the lateral resolution provided by optical focusing is degraded by only 14% when zf=1.1 transport mean free path (l t'), compared with the case of no scattering. When zf=1.7 lt', the fluence at 50 μm radial distance away from the focal point is 93% of that at the focal point, which shows optical focusing is very weak at this depth. The method to simulate the PSF of PAM can be used in the future to optimize parameters so as to improve the system performance.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。