Validity and reliability of assessing diaphragmatic mobility by area on X-rays of healthy subjects

健康受试者X光片上按区域评估膈肌活动度的有效性和可靠性

阅读:4
作者:Aline Pedrini, Márcia Aparecida Gonçalves, Bruna Estima Leal, Michelle Gonçalves de Souza Tavares, Wellington Pereira Yamaguti, David Luiz Góes, Elaine Paulin

Conclusions

The DMarea method proved to be a valid, reliable measure of diaphragmatic mobility.

Methods

We evaluated anthropometric parameters, pulmonary function, and diaphragmatic mobility in 43 participants. Two observers (rater A and rater B) determined diaphragmatic mobility at two time points. We used Pearson's correlation coefficient to evaluate the correlation between DMarea and the assessment of diaphragmatic mobility by distance (DMdist). To evaluate intra- and inter-rater reliability, we used the intraclass correlation coefficient (ICC [2,1]), 95% CI, and Bland-Altman analysis.

Objective

To investigate the concurrent validity, as well as the intra- and inter-rater reliability, of assessing diaphragmatic mobility by area (DMarea) on chest X-rays of healthy adults.

Results

A significant correlation was found between the DMarea and DMdist methods (r = 0.743; p < 0.0001). For DMarea, the intra-rater reliability was found to be quite high for the right hemidiaphragm (RHD)-ICC (2,1) = 0.92 (95% CI: 0.86-0.95) for rater A and ICC (2,1) = 0.90 (95% CI: 0.84-0.94) for rater B-and the left hemidiaphragm (LHD)-ICC (2,1) = 0.96 (95% CI: 0.93-0.97) for rater A and ICC (2,1) = 0.91 (95% CI: 0.81-0.95) for rater B-(p < 0.0001 for all). Also for DMarea, the inter-rater reliability was found to be quite high for the first and second evaluations of the RHD-ICC (2,1) = 0.99 (95% CI: 0.98-0.99) and ICC (2,1) = 0.95 (95% CI: 0.86-0.97), respectively-and the LHD-ICC (2,1) = 0.99 (95% CI: 0.98-0.99) and ICC (2,1) = 0.94 (95% CI: 0.87-0.97)-(p < 0.0001 for both). The Bland-Altman analysis showed good agreement between the mobility of the RHD and that of the LHD. Conclusions: The DMarea method proved to be a valid, reliable measure of diaphragmatic mobility.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。