Tackling SNR at low-field: a review of hardware approaches for point-of-care systems

解决低场信噪比问题:即时检测系统硬件方法综述

阅读:1

Abstract

OBJECTIVE: To review the major hardware components of low-field point-of-care MRI systems which affect the overall sensitivity. METHODS: Designs for the following components are reviewed and analyzed: magnet, RF coils, transmit/receive switches, preamplifiers, data acquisition system, and methods for grounding and mitigating electromagnetic interference. RESULTS: High homogeneity magnets can be produced in a variety of different designs including C- and H-shaped as well as Halbach arrays. Using Litz wire for RF coil designs enables unloaded Q values of ~ 400 to be reached, with body loss representing about 35% of the total system resistance. There are a number of different schemes to tackle issues arising from the low coil bandwidth with respect to the imaging bandwidth. Finally, the effects of good RF shielding, proper electrical grounding, and effective electromagnetic interference reduction can lead to substantial increases in image signal-to-noise ratio. DISCUSSION: There are many different magnet and RF coil designs in the literature, and to enable meaningful comparisons and optimizations to be performed it would be very helpful to determine a standardized set of sensitivity measures, irrespective of design.