Non-contact infrared thermometers compared with current approaches in primary care for children aged 5 years and under: a method comparison study

非接触式红外体温计与目前初级保健中针对5岁及以下儿童的测量方法的比较:一项方法比较研究

阅读:1

Abstract

BACKGROUND: Current options for temperature measurement in children presenting to primary care include either electronic axillary or infrared tympanic thermometers. Non-contact infrared thermometers could reduce both the distress of the child and the risk of cross-infection. OBJECTIVES: The objective of this study was to compare the use of non-contact thermometers with the use of electronic axillary and infrared tympanic thermometers in children presenting to primary care. DESIGN: Method comparison study with a nested qualitative study. SETTING: Primary care in Oxfordshire. PARTICIPANTS: Children aged ≤ 5 years attending with an acute illness. INTERVENTIONS: Two types of non-contact infrared thermometers [i.e. Thermofocus (Tecnimed, Varese, Italy) and Firhealth (Firhealth, Shenzhen, China)] were compared with an electronic axillary thermometer and an infrared tympanic thermometer. MAIN OUTCOME MEASURES: The primary outcome was agreement between the Thermofocus non-contact infrared thermometer and the axillary thermometer. Secondary outcomes included agreement between all other sets of thermometers, diagnostic accuracy for detecting fever, parental and child ratings of acceptability and discomfort, and themes arising from our qualitative interviews with parents. RESULTS: A total of 401 children (203 boys) were recruited, with a median age of 1.6 years (interquartile range 0.79-3.38 years). The readings of the Thermofocus non-contact infrared thermometer differed from those of the axillary thermometer by -0.14 °C (95% confidence interval -0.21 to -0.06 °C) on average with the lower limit of agreement being -1.57 °C (95% confidence interval -1.69 to -1.44 °C) and the upper limit being 1.29 °C (95% confidence interval 1.16 to 1.42 °C). The readings of the Firhealth non-contact infrared thermometer differed from those of the axillary thermometer by -0.16 °C (95% confidence interval -0.23 to -0.09 °C) on average, with the lower limit of agreement being -1.54 °C (95% confidence interval -1.66 to -1.41 °C) and the upper limit being 1.22 °C (95% confidence interval 1.10 to 1.34 °C). The difference between the first and second readings of the Thermofocus was -0.04 °C (95% confidence interval -0.07 to -0.01 °C); the lower limit was -0.56 °C (95% confidence interval -0.60 to -0.51 °C) and the upper limit was 0.47 °C (95% confidence interval 0.43 to 0.52 °C). The difference between the first and second readings of the Firhealth thermometer was 0.01 °C (95% confidence interval -0.02 to 0.04 °C); the lower limit was -0.60 °C (95% confidence interval -0.65 to -0.54 °C) and the upper limit was 0.61 °C (95% confidence interval 0.56 to 0.67 °C). Sensitivity and specificity for the Thermofocus non-contact infrared thermometer were 66.7% (95% confidence interval 38.4% to 88.2%) and 98.0% (95% confidence interval 96.0% to 99.2%), respectively. For the Firhealth non-contact infrared thermometer, sensitivity was 12.5% (95% confidence interval 1.6% to 38.3%) and specificity was 99.4% (95% confidence interval 98.0% to 99.9%). The majority of parents found all methods to be acceptable, although discomfort ratings were highest for the axillary thermometer. The non-contact thermometers required fewer readings than the comparator thermometers. LIMITATIONS: A method comparison study does not compare new methods against a reference standard, which in this case would be central thermometry requiring the placement of a central line, which is not feasible or acceptable in primary care. Electronic axillary and infrared tympanic thermometers have been found to have moderate agreement themselves with central temperature measurements. CONCLUSIONS: The 95% limits of agreement are > 1 °C for both non-contact infrared thermometers compared with electronic axillary and infrared tympanic thermometers, which could affect clinical decision-making. Sensitivity for fever was low to moderate for both non-contact thermometers. FUTURE WORK: Better methods for peripheral temperature measurement that agree well with central thermometry are needed. TRIAL REGISTRATION: Current Controlled Trials ISRCTN15413321. FUNDING: This project was funded by the National Institute for Health Research (NIHR) Health Technology Assessment programme and will be published in full in Health Technology Assessment; Vol. 24, No. 53. See the NIHR Journals Library website for further project information.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。