Gamma radiation detector selection for CT scanner

CT扫描仪的伽马射线探测器选择

阅读:1

Abstract

Three types of gamma radiation detectors associated with distributed electronics namely, NaI (Tl), HPGe and LaBr(3)(Ce) are compared primarily focusing on electronic noise and scattering noise. Additionally, detectors of same make, material, size and electronics are also compared. A methodology is proposed to select the most suitable detector for computed tomography (CT) among the available options. Standard deviation parameter is employed to estimate electronic noise without performing CT experiment. Kanpur theorem-1(KT-1) is used to estimate the scattering noise quantitatively after verifying its sensitivity to scattering noise. The impact of scattering noise on CT profiles is evaluated using dice similarity dice coefficient. A good resemblance between KT-1 and dice coefficient is observed. A maximum difference of 56% in scattering noise is observed when five detectors used simultaneously instead of single detector whereas a discrepancy of 85% is observed between different types of radiation detectors. As far as ease of handling, operational and capital cost is concern one has to compromise minimum 12% of accuracy in CT reconstruction if NaI (Tl) detector is used with respect to best alternative available. The proposed methodology can be applied to measurement that require minimal scattering interference data other than CT experiments. The manufacturer can add noise level of detector as a characteristic parameter in the data sheet.