Implementation of an Environmental Cleaning Protocol in Hospital Critical Areas Using a UV-C Disinfection Robot

在医院关键区域实施环境清洁规程,采用紫外线消毒机器人

阅读:1

Abstract

Improving the cleaning and disinfection of high-touch surfaces is one of the core components of reducing healthcare-associated infections. The effectiveness of an enhanced protocol applying UV-C irradiation for terminal room disinfection between two successive patients was evaluated. Twenty high-touch surfaces in different critical areas were sampled according to ISO 14698-1, both immediately pre- and post-cleaning and disinfection standard operating protocol (SOP) and after UV-C disinfection (160 sampling sites in each condition, 480 in total). Dosimeters were applied at the sites to assess the dose emitted. A total of 64.3% (103/160) of the sampling sites tested after SOP were positive, whereas only 17.5% (28/160) were positive after UV-C. According to the national hygienic standards for health-care setting, 9.3% (15/160) resulted in being non-compliant after SOP and only 1.2% (2/160) were non-compliant after UV-C disinfection. Operation theaters was the setting that resulted in being less compliant with the standard limit (≤15 colony-forming unit/24 cm(2)) after SOP (12%, 14/120 sampling sites) and where the UV-C treatment showed the highest effectiveness (1.6%, 2/120). The addition of UV-C disinfection to the standard cleaning and disinfection procedure had effective results in reducing hygiene failures.