Differences in image density adjustment parameters on the image matching accuracy of a floor-mounted kV X-ray image-guided radiation therapy system

落地式千伏X射线图像引导放射治疗系统的图像密度调整参数差异对图像匹配精度的影响

阅读:1

Abstract

This study aimed to investigate the effect of two different image density adjustment parameters on the results of image matching at six degrees of freedom using radiographic images generated by the ExacTrac X-ray system in brain stereotactic radiosurgery (SRS). This study comprised 32 patients who underwent brain SRS at our hospital from January 2020 to December 2020. In this study, (1) the default parameter (an image density parameter between "tissue" and "bone") was an image density parameter for digitally reconstructed radiograph (DRR) generation used at many facilities, and (2) the bone parameter was the steepest contrast parameter used at our hospital. Of the 32 patients, 24 (75%) had a couch angle of 0.5 mm or more in the translational direction or 0.5° or more in the rotational direction, and 10 (31%) had a couch angle of 1.0 mm or more in the translational direction or 1.0° or more in the rotational direction. Among the 131 cases of all couch angles, 46 (35%) cases had a translational direction of 0.5 mm or more or a rotational direction of 0.5° or more, and 15 (11%) had a translational direction of 1.0 mm or more or a rotational direction of 1.0° or more. The results of this study indicate the usefulness of using appropriate DRR parameters for each case, rather than using the default settings. The use of appropriate DRR parameters can lead to accurate position matching results, leading to fewer image-guided radiation therapy shots and a lower imaging dose.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。