A finite element analysis of a low-profile femoral neck system of screws in sleeves in a vertical femoral neck fracture model

对垂直股骨颈骨折模型中低轮廓套筒螺钉股骨颈系统进行有限元分析

阅读:1

Abstract

BACKGROUND: Femoral neck system (FNS) has exhibited some drawbacks, such as non-fit of the plate with the lateral femoral cortex, postoperative pain, and the potential risk of subtrochanteric fractures. We have developed a low-profile FNS system that addresses some compatibility issues in FNS. In this study, we conducted finite element analysis on the 1-hole FNS (1 H-FNS), 2-holes FNS (2 H-FNS), and low-profile FNS (LP-FNS) and compared their biomechanical performance. METHODS: After the mesh convergence analysis, we established three groups of 1 H-FNS, 2 H-FNS, and LP-FNS. The interfragmentary gap, sliding distance, shear stress, and compressive stress and the bone-implant interface compression stress, stiffness, and displacement were determined under the neutral, flexion, or extension conditions of the hip joint, respectively. The stress and displacement of the femur after the implant removal were also investigated. RESULTS: (1) There were no obvious differences among the three FNS groups in terms of the IFM distance. However, the LP-FNS group showed less rotational angle compared with conventional FNS (neutral: 1 H-FNS, -61.64%; 2 H-FNS, -45.40%). Also, the maximum bone-implant interface compression stress was obviously decreased under the neutral, flexion, or extension conditions of the hip joint (1 H-FNS: -6.47%, -20.59%, or -4.49%; 2 H-FNS: -3.11%, 16.70%, or -7.03%; respectively). (2) After the implant removal, there was no notable difference in the maximum displacement between the three groups, but the maximum von Mises stress displayed a notable difference between LP-FNS and 1 H-FNS groups (-15.27%) except for the difference between LP-FNS and 2 H-FNS groups (-4.57%). CONCLUSIONS: The LP-FNS may not only provide the same biomechanical stabilities as the 1 H-FNS and 2 H-FNS, but also have more advantages in rotational resistance especially under the neutral condition of the hip joint, in the bone-implant interface compression stress, and after the implant removal. In addition, the 1 H-FNS and 2 H-FNS have similar biomechanical stabilities except for the maximum von Mises stress after the implant removal. The femur after the LP-FNS removal not only is subjected to relatively little stress but also minimizes stress concentration areas.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。