Performance of BinaxNOW G6PD deficiency point-of-care diagnostic in P. vivax-infected subjects

BinaxNOW G6PD 缺乏症即时诊断在间日疟原虫感染者中的性能

阅读:4

Abstract

Accurate diagnosis of glucose-6-phosphate dehydrogenase (G6PD) deficiency is required to avoid the risk of acute hemolysis associated with 8-aminoquinoline treatment. The performance of the BinaxNOW G6PD test compared with the quantitative spectrophotometric analysis of G6PD activity was assessed in 356 Plasmodium vivax-infected subjects in Brazil, Peru, Thailand, and India. In the quantitative assay, the median G6PD activity was 8.81 U/g hemoglobin (range = 0.05-20.19), with 11 (3%) subjects identified as deficient. Sensitivity of the BinaxNOW G6PD to detect deficient subjects was 54.5% (6 of 11), and specificity was 100% (345 of 345). Room temperatures inadvertently falling outside the range required to perform the rapid test (18-25°C) together with subtlety of color change and insufficient training could partially explain the low sensitivity found. Ensuring safe use of 8-aminoquinolines depends on additional development of simple, highly sensitive G6PD deficiency diagnostic tests suitable for routine use in malaria-endemic areas.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。