Temperature and the field stability of a dengue rapid diagnostic test in the tropics

温度与登革热快速诊断测试在热带地区的现场稳定性

阅读:1

Abstract

The global incidence of dengue has increased significantly in recent decades, resulting in a large public health burden in tropical and subtropical countries. Dengue rapid diagnostic tests (RDTs) can provide accurate, rapid accessible diagnosis for patient management and may be easily used by health workers in rural areas. However, in dengue-endemic areas, ambient temperatures are often higher than manufacturer's recommendation. We therefore evaluated the effect of high temperature over time on the performance of one commonly used dengue RDT, the Standard Diagnostics Bioline Dengue Duo. RDTs were kept in five different conditions (at 4°C, 35°C, 45°C, 60°C, and at fluctuant ambient temperatures in a free-standing hut) for between 2 days and 2 years in the Lao People's Democratic Republic (PDR). RDTs were tested with four control sera (negative, dengue nonstructural protein 1 [NS1], anti-dengue immunoglobulin [Ig] M, and anti-dengue IgG positive). The RDTs had 100% consistency over the 2-year study, despite high temperatures, including in the hut in which temperatures exceeded the manufacturer's recommendations for 29% of time points. These data suggest that the diagnostic accuracy of the SD Bioline Dengue Duo RDT remains stable even after long-term storage at high temperatures. Therefore, use at such ambient temperatures in tropical areas should not jeopardize the dengue diagnostic outcome.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。