Evaluating clinical near-infrared surgical camera systems with a view to optimizing operator and computational signal analysis

评估临床近红外手术摄像系统,旨在优化操作者和计算信号分析

阅读:1

Abstract

SIGNIFICANCE: As clinical evidence on the colorectal application of indocyanine green (ICG) perfusion angiography accrues, there is also interest in computerizing decision support. However, user interpretation and software development may be impacted by system factors affecting the displayed near-infrared (NIR) signal. AIM: We aim to assess the impact of camera positioning on the displayed NIR signal across different open and laparoscopic camera systems. APPROACH: The effects of distance, movement, and target location (center versus periphery) on the displayed fluorescence signal of different systems were measured under electromagnetic stereotactic guidance from an ICG-albumin model and in vivo during surgery. RESULTS: Systems displayed distinct fluorescence performances with variance apparent with scope optical lens configuration (0 deg versus 30 deg), movement, target positioning, and distance. Laparoscopic system readings fitted inverse square function distance-intensity curves with one device and demonstrated a direction dependent sigmoid curve. Laparoscopic cameras presented central targets as brighter than peripheral ones, and laparoscopes with angled optical lens configurations had a diminished field of view. One handheld open system also showed a distance-intensity relationship, whereas the other maintained a consistent signal despite distance, but both presented peripheral targets brighter than central ones. CONCLUSIONS: Optimal clinical use and signal computational development requires detailed appreciation of system behaviors.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。