Gaussian process regression for ultrasound scanline interpolation

用于超声扫描线插值的高斯过程回归

阅读:1

Abstract

Purpose: In ultrasound imaging, interpolation is a key step in converting scanline data to brightness-mode (B-mode) images. Conventional methods, such as bilinear interpolation, do not fully capture the spatial dependence between data points, which leads to deviations from the underlying probability distribution at the interpolation points. Approach: We propose Gaussian process ( GP ) regression as an improved method for ultrasound scanline interpolation. Using ultrasound scanlines acquired from two different ultrasound scanners during in vivo trials, we compare the scanline conversion accuracy of three standard interpolation methods with that of GP regression, measuring the peak signal-to-noise ratio (PSNR) and mean absolute error (MAE) for each method. Results: The PSNR and MAE scores show that GP regression leads to more accurate scanline conversion compared to the nearest neighbor, bilinear, and cubic spline interpolation methods, for both datasets. Furthermore, limiting the interpolation window size of GP regression to 15 reduces computation time with minimal to no reduction in PSNR. Conclusions: GP regression quantitatively leads to more accurate scanline conversion and provides uncertainty estimates at each of the interpolation points. Our windowing method reduces the computational cost of using GP regression for scanline conversion.