Co-axial acoustic-based optical coherence vibrometry probe for the quantification of resonance frequency modes in ocular tissue

用于量化眼组织共振频率模式的同轴声学光学相干振动测量探头

阅读:2

Abstract

We present a co-axial acoustic-based optical coherence vibrometry probe (CoA-OCV) for vibro-acoustic resonance quantification in biological tissues. Sample vibrations were stimulated via a loudspeaker, and pre-compensation was used to calibrate the acoustic spectrum. Sample vibrations were measured via phase-sensitive swept-source optical coherence tomography (OCT). Resonance frequencies of corneal phantoms were measured at varying intraocular pressures (IOP), and dependencies on Young´s Modulus (E), phantom thickness and IOP were observed. Cycling IOP revealed hysteresis. For E = 0.3 MPa, resonance frequencies increased with IOP at a rate of 3.9, 3.7 and 3.5 Hz/mmHg for varied thicknesses and 1.7, 2.5 and 2.8 Hz/mmHg for E = 0.16 MPa. Resonance frequencies increased with thickness at a rate of 0.25 Hz/µm for E = 0.3 MPa, and 0.40 Hz/µm for E = 0.16 MPa. E showed the most predominant impact in the shift of the resonance frequencies. Full width at half maximum (FWHM) of the resonance modes increased with increasing thickness and decreased with increasing E. Only thickness and E contributed to the variance of FWHM. In rabbit corneas, resonance frequencies of 360-460 Hz were observed. The results of the current study demonstrate the feasibility of CoA-OCV for use in future OCT-V studies.