Low-threshold, high-resolution, chronically stable intracortical microstimulation by ultraflexible electrodes

低阈值、高分辨率、通过超柔性电极进行长期稳定的皮层内微刺激

阅读:5
作者:Roy Lycke, Robin Kim, Pavlo Zolotavin, Jon Montes, Yingchu Sun, Aron Koszeghy, Esra Altun, Brian Noble, Rongkang Yin, Fei He, Nelson Totah, Chong Xie, Lan Luan

Abstract

Intracortical microstimulation (ICMS) enables applications ranging from neuroprosthetics to causal circuit manipulations. However, the resolution, efficacy, and chronic stability of neuromodulation are often compromised by adverse tissue responses to the indwelling electrodes. Here we engineer ultraflexible stim-nanoelectronic threads (StimNETs) and demonstrate low activation threshold, high resolution, and chronically stable ICMS in awake, behaving mouse models. In vivo two-photon imaging reveals that StimNETs remain seamlessly integrated with the nervous tissue throughout chronic stimulation periods and elicit stable, focal neuronal activation at low currents of 2 μA. Importantly, StimNETs evoke longitudinally stable behavioral responses for over 8 months at a markedly low charge injection of 0.25 nC/phase. Quantified histological analyses show that chronic ICMS by StimNETs induces no neuronal degeneration or glial scarring. These results suggest that tissue-integrated electrodes provide a path for robust, long-lasting, spatially selective neuromodulation at low currents, which lessens risk of tissue damage or exacerbation of off-target side effects.