5G Service and Pacemakers/Implantable Defibrillators: What Is the Actual Risk?

5G 服务与心脏起搏器/植入式除颤器:实际风险是什么?

阅读:1

Abstract

Rapidly growing worldwide, 5G service is expected to deeply change the way we communicate, connect and share data. It encompasses the whole spectrum of new technology, infrastructure and mobile connectivity, and will touch not only every sector in the industry, but also many aspects of our everyday life. Although the compliance with international regulations provides reasonable protection to public health and safety, there might be specific issues not fully covered by the current technical standards. Among the aspects that shall be carefully considered, there is the potential interference that can be induced on medical devices, and in particular on implantable medical devices that are critical for the patient's life, such as pacemakers and implantable defibrillators. This study aims to assess the actual risk that 5G communication systems pose to pacemakers and implantable defibrillators. The setup proposed by the ISO 14117 standard was adapted to include 5G characteristic frequencies of 700 MHz and 3.6 GHz. A total number of 384 tests were conducted. Among them, 43 EMI events were observed. Collected results reveal that RF hand-held transmitters operating in these two frequency bands do not pose additional risk compared to pre-5G bands and that the safety distance of 15 cm typically indicted by the PM/ICD manufacturer is still able to guarantee the patient's safety.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。