Comparing Equil patch versus traditional catheter insulin pump in type 2 diabetes using continuous glucose monitoring metrics and profiles

利用连续血糖监测指标和曲线,比较 Equil 贴片与传统导管胰岛素泵在 2 型糖尿病中的应用。

阅读:1

Abstract

AIMS: It is not clear whether there are differences in glycemic control between the Equil patch and the MMT-712 insulin pump. Our objective was to compare two types of insulin pumps in the treatment of type 2 diabetes mellitus (T2DM), using continuous glucose monitoring (CGM) metrics and profiles. METHODS: This was a randomized case-crossover clinical trial. Participants were hospitalized and randomly allocated to two groups and underwent two types of insulin pump treatments (group A: Equil patch-Medtronic MMT-712 insulin pump; group B: Medtronic MMT-712-Equil patch insulin pump) separated by a 1-day washout period. Glycemic control was achieved after 7-8 days of insulin pump therapy. Each patient received CGM for 5 consecutive days (from day 1 to day 5). On day 3 of CGM performance, the Equil patch insulin pump treatment was switched to Medtronic MMT-712 insulin pump treatment at the same basal and bolus insulin doses or vice versa. CGM metrics and profiles including glycemic variability (GV), time in range (TIR, 3.9-10.0 mmol/L), time below range (TBR, <3.9 mmol/L), time above range (TAR, >10.0 mmol/L), and postprandial glucose excursions, as well as incidence of hypoglycemia. RESULTS: Forty-six T2DM patients completed the study. There was no significant difference in parameters of daily GV and postprandial glucose excursions between the Equil patch insulin pump treatment and the Medtronic insulin pump treatment. Similarly, there was no between-treatment difference in TIR, TBR, and TAR, as well as the incidence of hypoglycemia. CONCLUSION: The Equil patch insulin pump was similar to the traditional MMT-712 insulin pump in terms of glycemic control. Equil patch insulin pump is a reliable tool for glycemic management of diabetes mellitus.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。