Determining urea levels in exhaled breath condensate with minimal preparation steps and classic LC-MS

采用最少的制备步骤和经典 LC-MS 法测定呼出气冷凝液中的尿素水平

阅读:8
作者:Masha Pitiranggon, Matthew S Perzanowski, Patrick L Kinney, Dongqun Xu, Steven N Chillrud, Beizhan Yan

Abstract

Exhaled breath condensate (EBC) provides a relatively easy, non-invasive method for measuring biomarkers of inflammation and oxidative stress in the airways. However, the levels of these biomarkers in EBC are influenced, not only by their levels in lung lining fluid but also by the volume of water vapor that also condenses during EBC collection. For this reason, the use of a biomarker of dilution has been recommended. Urea has been proposed and utilized as a promising dilution biomarker due to its even distribution throughout the body and relatively low volatility. Current EBC urea analytical methods either are not sensitive enough, necessitating large volumes of EBC, or are labor intensive, requiring a derivatization step or other pretreatment. We report here a straightforward and reliable LC-MS approach that we developed that does not require derivatization or large sample volume (∼36 µL). An Acclaim mixed-mode hydrophilic interaction chromatography column was selected because it can produce good peak symmetry and efficiently separate urea from other polar and nonpolar compounds. To achieve a high recovery rate, a slow and incomplete evaporation method was used followed by a solvent-phase exchange. Among EBC samples collected from 28 children, urea levels were found to be highly variable, with a relative standard deviation of 234%, suggesting high variability in dilution of the lung lining fluid component of EBC. The limit of detection was found to be 0.036 µg/mL.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。