BACKGROUND: Dog Bone™ button fixation is frequently used to treat acromioclavicular joint (ACJ) dislocation. However, various studies have reported complications after fixation. OBJECTIVE: To investigate the effect of the coracoid bone tunnel location on the treatment of ACJ dislocation through single-tunnel coracoclavicular (CC) ligament fixation with the Dog Bone™ button. METHODS: Six cadaveric shoulders were used. Each specimen was subjected to five testing conditions in the following order: (1) normal ACJ (Gn); (2) acromioclavicular and CC ligaments were removed (G0); (3) CC ligament reconstruction was performed using the Dog Bone™ technique, and the coracoid bone tunnel was at the center of the coracoid base (G1); (4) reconstruction was performed at 5 mm distal from the G1 site, along the axis of the coracoid (G2); (5) reconstruction was performed at 10 mm distal from the G1 site, along the axis of the coracoid (G3). The angles of pronation and supination of the clavicle under the same load (30 N) were measured. Next, a finite element (FE) model was created using computed tomography (CT) images of the normal shoulder. Model 1 (M1), model 2 (M2), and model 3 (M3) correspond to G1, G2, and G3, respectively. A force of 70 N was applied as a vertical upward load to the distal clavicle. Subsequently, the von Mises stress, the strain LE along the FiberWire, and the displacement nephogram of the three models were obtained. RESULTS: After single-tunnel CC ligament fixation using the Dog Bone™ technique, the clavicle in the G2 group (20.50 (19.50, 21.25) °, 20.00 (18.75, 21.25) °) had the best rotational stability. The peak von Mises stress, the strain LE along the FiberWire, and the maximum displacement were smaller in M2 than in M1 and M3. CONCLUSIONS: When the coracoid bone tunnel was located 5 mm anterior to the center of the coracoid base (along the axis of the coracoid), the clavicle showed greater rotational stability.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。