Assessing component orientation of total hip arthroplasty using the low-dose bi-planar radiographs

利用低剂量双平面X线片评估全髋关节置换术的组件方向

阅读:1

Abstract

BACKGROUND: Three-dimensional computed tomography (3D CT) reconstruction is the reference standard for measuring component orientation. However, functional cup orientation in standing position is preferable compared with supine position. The low-dose bi-planar radiographs can be used to analyze standing cup component orientation. We aimed to assess the validity and reliability of the component orientation using the low-dose bi-planar radiographs compared with the 3D CT reconstruction, and explore the differences between the functional cup orientation in standing radiographs and supine CT scans. METHODS: A retrospective study, including 44 patients (50 hips) with total hip arthroplasty (THA), was conducted. CT scans were taken 1 week after surgery and the low-dose bi-planar radiographs were taken in the follow-up 6 weeks later. Component orientation measurement was performed using the anterior pelvic plane and the radiographic coronal plane as reference, respectively. RESULTS: The study showed no significant difference in cup anteversion (p = 0.160), cup inclination (p = 0.486), and stem anteversion (p = 0.219) measured by the low-dose bi-planar radiographs and 3D reconstruction. The differences calculated by the Bland-Altman analysis ranged from - 0.4° to 0.6° for the three measured angles. However, the mean absolute error was 4.76 ± 1.07° for functional anteversion (p = 0.035) and 4.02 ± 1.08° for functional inclination (p = 0.030) measured by the bi-planar radiographs and supine CT scans. CONCLUSIONS: The low-dose bi-planar radiographs are the same reliable and accurate as 3D CT reconstruction to assess post-THA patients' component orientation, while providing more valuable functional component orientation than supine CT scans.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。