Liner dissociation in a large-diameter ceramic-bearing acetabular component: a report of five cases

大直径陶瓷轴承髋臼假体内衬脱位:五例报告

阅读:3

Abstract

BACKGROUND: Ceramic-on-ceramic (CoC) bearings for total hip arthroplasty (THA) have been offering very favorable results and survivorship since their introduction. In order to increase range of movement (ROM) and decrease dislocation rates, some manufacturers have introduced larger diameter head (LDH) CoC bearings. This has been achieved with the use of preassembled cup designs, in which the ceramic liner is already fitted into the metal backing and implanted as a monoblock component by the surgeon. In this report we present data from a series of 5 patients with ceramic liner dissociation from a monoblock cup. CASE PRESENTATION: All cases were overweight men with acetabular components of 56 or 58 mm. After a mean of 5.5 (range, 3.5-6.7) years, all patients reported sudden pain and audible noise when performing activities of daily living. Liner displacement was suspected on plain radiographs and confirmed by Ct-scan. Pneumarthrosis was present in all cases. Taper modular junction wear and corrosion signs were observed in the four revised patients. CONCLUSION: Although one of our case is still treated conservatively, implant revision is probably inevitable. Further LDH CoC implant design should take in consideration this potential complication by avoiding bearing diameters over 40mm and/or improving locking mechanism or by providing a real monoblock acetabular implant.