Bilateral neck fracture in bimodular femoral stem after primary total hip arthroplasty: a case report

初次全髋关节置换术后双模块股骨柄双侧颈骨折:病例报告

阅读:1

Abstract

BACKGROUND: Bi-modular stems were introduced in primary total hip arthroplasty (THA) to enable better control of the femoral offset, leg length, and hip stability. Despite numerous reports on modular femoral neck fractures, some designs are still marketed worldwide. While the risk factors for the sudden failure are multifactorial and mostly known, the timing of this new THA complication is not predictable by any means. CASE PRESENTATION: In this report, the literature regarding one of the most popular bi-modular stems with specific neck-stem coupling (oval Morse taper) is reviewed and illustrated with a case of bilateral modular neck fracture in a patient with idiopathic aseptic necrosis of femoral heads treated with primary bi-modular THA. Because of bilateral modular femoral neck fracture, which occurred 3 years on the left side and 20 years after implantation on the right side, the patient required a total of 6 revisions and 208 days of hospitalized care. CONCLUSION: To our knowledge, this is the first report of bilateral modular neck fracture in a single patient. Even though the same surgeon performed both operations and used the same neck length and orientation, fractures occurred with a 17-year time difference after implantation. This shows that we cannot predict with certainty when a fracture might occur. Orthopaedic surgeons should use bi-modular stem designs for primary THA very cautiously.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。