An optimization method for implantation parameters of individualized TKA tibial prosthesis based on finite element analysis and orthogonal experimental design

基于有限元分析和正交试验设计的个体化全膝关节置换术胫骨假体植入参数优化方法

阅读:1

Abstract

BACKGROUND: Individualized and accurate implantation of a tibial prosthesis during total knee arthroplasty (TKA) can assist in uniformly distributing the load and reducing the polyethylene wear to obtain a long-term prosthetic survival rate, but individualized and accurate implantation of a tibial prosthesis during TKA remains challenging. The purpose of this study was to optimize and individualize the positioning parameters of a tibial prosthesis to improve its accurate implantation using a new method of finite element analysis in combination with orthogonal experimental design. METHODS: Ten finite element models of TKA knee joint were developed to optimize the implantation parameters (varus angle, posterior slope angle, and external rotation angle) of tibial prosthesis to reduce the peak value of the contact pressure on the polyethylene liner according to the method of finite element analysis in combination with orthogonal experimental design. The influence of implantation parameters on the peak value of the contact pressure on the polyethylene liner was evaluated based on a range analysis in orthogonal experimental design. RESULTS: The optimal implantation parameters for tibial prosthesis included 0° varus, 1° posterior slope, and 4° external rotation. Under these conditions, the peak value of the contact pressure on the polyethylene liner remained the smallest (16.37 MPa). Among the three parameters that affect the peak value of the contact pressure, the varus angle had the greatest effect (range = 6.70), followed by the posterior slope angle (range = 2.36), and the external rotation angle (range = 2.15). CONCLUSIONS: The optimization method based on finite element analysis and orthogonal experimental design can guide the accurate implantation of the tibial prosthesis, reducing the peak value of the contact pressure on the polyethylene liner. This method provides new insights into the TKA preoperative plan and biomechanical decision-making for accurately implanting TKA prosthesis.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。