Introduction of a guide based on a femoral neck section for fixation with multiple screws: a cadaveric study

基于股骨颈截面的多螺钉固定导向器的引入:一项尸体研究

阅读:1

Abstract

BACKGROUND: The design and application of assisted instrumentation for internal fixation of femoral neck fractures with multiple screws are still evolving. A novel guide based on a femoral neck section was designed to improve the accuracy of screw placement, and its efficacy was evaluated. METHODS: A guide based on a femoral neck section was designed for assisted fixation of femoral neck fractures with multiple screws. Femoral specimens from 10 adults (20 femurs) underwent assisted internal fixation for a femoral neck fracture with 3 cannulated screws using the new guide technique or conventional technique. The accuracy of screw orientation and entry point, the accuracy of optimal screw positioning, and drilling attempts, operative time, and fluoroscopy time were recorded. RESULTS: Among all 20 specimens, 60 screws were inserted successfully. Screw parallelism, operative time, and fluoroscopy time showed no statistical difference between the new guide technique and conventional technique (P > 0.05). The accuracy of optimal screw positioning was determined by the contained screw area ratio, distance between screws, distance from the centre of the femoral neck section, distance between screws and the femoral neck cortex, and Drilling attempts were statistical significantly better (data in the first three were larger and in the latter two was smaller) with the new guide technique, than with conventional technique (P < 0.05). CONCLUSIONS: This new, two-dimensional, fluoroscopy-assisted, percutaneous guide technique enables accurate and optimal screw positioning in internal fixation of femoral neck fractures, compared with conventional technique.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。