Thermal ablation a comparison of thermal dose required for radiofrequency-, microwave-, and laser-induced coagulation in an ex vivo bovine liver model

热消融:离体牛肝模型中射频、微波和激光诱导凝固所需热剂量的比较

阅读:1

Abstract

RATIONALE AND OBJECTIVES: To compare thermal dosimetry metrics for specified diameters of coagulation achieved using three different ablation energy sources. MATERIALS AND METHODS: 204 ablations measuring 20, 30, or 40 +/- 2 mm were created in an ex-vivo bovine liver model using 1) 2.5 cm cluster RF electrodes (n = 114), 2) 3 cm microwave antennas (n = 45), and 3) 3 cm laser diffusing fibers (n = 45). Continuous temperature monitoring was performed 5-20 mm from the applicators to calculate: a) the area under the curve (AUC), b) cumulative equivalent minutes at 43 degrees C (CEM43), and c) Arrhenius damage integral (Omega) for the critical ablation margin (DOC), with results compared by multivariate analysis of variance and regression analysis. RESULTS: The end temperatures at the margin of coagulation varied, and was lowest for the RF cluster electrode (33-58 degrees C), higher for laser (52-72 degrees C), and covered the widest range for microwave (42-95 degrees C). These end temperatures correlated with applied energy, as linear functions (r(2) = 0.74-0.96). The total heat needed to achieve ablation (AUC) varied with applied energy and coagulation diameter as negative exponential (RF and laser) or negative power (microwave) functions (r(2) = 0.82-0.98). Similarly, CEM43 values varied exponentially with energy and distance (r(2) = 0.52-0.76) over a wide range of values (10(12)). Likewise, Omega varied not only based upon energy source and DOC, but also as a positive linear correlation to applied energy and with sigmoid correlation to duration of ablation (r(2) = 0.85-0.97). CONCLUSION: Our study demonstrates that the thermal dosimetry of ablation is not based solely on a fixed end temperature at the margin of the coagulation zone. Thermal dosimetry is not constant, but dependent on the type and amount of energy applied and distance suggesting the need to take into account the rate of heat transfer for ablation dosimetry.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。