Trace lead measurement and online removal of matrix interference in geosamples by ion-exchange coupled with flow injection and hydride generation atomic fluorescence spectrometry

利用离子交换-流动注射-氢化物发生原子荧光光谱法测定地质样品中的痕量铅含量并在线去除基质干扰

阅读:1

Abstract

A flow injection method has been developed for the direct determination of free available Pb(II). The method is based on the chemical sorption of Pb(II), from pH7 solutions, on a column packed of chelating resin. The retained complex was afterwards eluted with hydrochloric acid followed by hydride generation with reduction by tetrahydroborate. The preconcentration system proposed in this paper allows the elimination of great part of the saline content in the sample. A thorough scrutiny was made for chemical variables and FI parameters. With a sampling volume of 10.5 mL, quantitative retention of Pb (II) was obtained, along with an enrichment factor of 40 and a sampling frequency of 15 h(-1). The detection limit, defined as 3 times the blank standard deviation (3sigma), was 0.0031 ngml(-1). The precision was characterized by an RSD value of 3.78% (at the 4 ng.ml(-1) level, n = 11). The developed method has been applied to the determination of trace Pb in three standard reference materials. Accuracy was assessed through comparing the results with the accepted values.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。