Quality control protocol for confocal systems

共聚焦系统的质量控制规程

阅读:1

Abstract

Quantitative analysis of confocal imaging experiments require more stringent quality control of instrument function than qualitative imaging. Unfortunately, there are no standard procedures for quality control that are uniformly implemented, and, in multi user facilities experimenters rarely have access to the QC information. This paper proposes an easy and very efficient protocol that could be performed at the beginning of each day, experiment or even slide. It takes only a few minutes to assess laser stability, stage stability, channel registration in 3 dimensions and flatness of field. The information may be used either to calibrate data or, in more severe cases to request servicing the instrument.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。