Analysis of the influence of surgical robot drilling parameters on the temperature of skull drilling based on Box-Behnken design

基于Box-Behnken设计的手术机器人钻孔参数对颅骨钻孔温度影响分析

阅读:1

Abstract

It is easy to cause thermal damage to the bone tissue when the surgical robot performs skull drilling to remove bone flaps, due to the large diameter of the drill bit, the large heat-generating area, and the long drilling time. Therefore, in order to reduce the thermal damage during the robot-assisted skull drilling process, the relationship between the drilling parameters and the drilling temperature during the skull drilling was studied in this paper. Firstly, a dynamic numerical simulation model of the skull drilling process was established by ABAQUS, and a temperature simulation plan for skull drilling was designed based on the Box-Behnken method. Then according to the simulation results, a quadratic regression model of drill diameter, feed rate, drill speed, and drilling temperature was established by using the multiple regression method. By analyzing the regression model, the influence of drilling parameters on the drilling temperature was clarified. Finally, the bone drilling experiment was carried out, and the error percentage was lower than 10.5% through the experiment to verify the reliability of the conclusion, and a safety strategy was proposed to ensure the safety of the surgical drilling process based on this experiment.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。