Necessity to critically review the automatic results of the Xpert Flu assay

有必要对Xpert Flu检测的自动结果进行严格审查

阅读:1

Abstract

While using the Xpert Flu assay we became aware of false-negative results. The study aimed to analyze the causes of these false-negative results. One hundred fifty-nine respiratory specimens were tested in the Xpert Flu assay and in multiplex reverse transcription-polymerase chain reactions (RT-PCRs) for respiratory viruses. Discordant specimens were tested in the Influenza A/B r-gene assay. One hundred fifty-two (96%) and 151 (95%) specimens yielded concordant results for influenza A and B, respectively. Fifteen specimens tested negative in the Xpert Flu assay and positive in a multiplex RT-PCR. Positive results were confirmed for 12 of these specimens in the Influenza A/B r-gene assay. Xpert Flu assay amplification curves and endpoints suggested that the false-negative results were mainly due to erroneous automatic result interpretation. We report false-negative results of the Xpert Flu assay due to erroneous automatic result interpretation. Careful analysis of amplification curves and endpoints is needed to avoid reporting of false-negative results.